Tehnične specifikacije SIRIUS® XHS-PWR

Specifikacije SIRIUS® XHS HybridADC

ADC
TypeHybrid ADC
High bandwidth mode16 bit SAR, up to 15 MS/s
Alias free mode24 bit Sigma Delta, up to 2 MS/s
Digital low pass filterSoftware selectable (off, alias free)
Analog HW filterAnalog 1 MHz 6th order (Butterworth)
Alias free filter0.5*fs cutoff, 0.45*fs pass band, 0.55*fs stop band to -100dB, zero phase distortion, available below 2MS/s

Specifikacija filtriranja SIRIUS® XHS HybridADC

Filter ModeHigh-bandwidth1 MHz LPFHigh-dynamic
Max. SR15 MS/s15 MS/s2 MS/s
SW settingsOff1 MHzAAF
Bandwidth (-3 dB)5 MHz1 MHz0.49 fs
Passband frequency (-0.1 dB)500 kHz500 kHz0.45 fs
Passband flatness± 0.1 dB± 0.1 dB± 0.1 dB
Alias-free bandwidth7.5 MHz0.45 fs
HW filter
Stopband frequency7.5 MHz0.55 fs
Stopband attenuation-100 dB-100 dB

Tehnične specifikacije SIRIUSi-XHS-PWR-1xHV-1xDCCT-1000A-UNI

SIRIUSi-XHS-PWR specificationsSIRIUSi-XHS-PWR-1xHV-1xDC-CT-1000A-UNI
Input typesVoltage, Current (single phase)
Voltage Mode
Ranges±2000 V, ±1000 V, ±400 V, ±200 V
ADC TypeHybrid ADC - alias free up to 2 MS/s, 16-bit up to 15 MS/s
Sampling RateSimultaneous 15 MS/s
FilteringAAF 1 MHz (6th order)
Analog bandwidth (-3 dB)5 MHz
Working voltage (continuous)1000 VDC or 1000 VAC
Measuring voltage (max.)±2000 V (see 1)
Input AccuracySignal frequencyAccuracy
DC±0.03 % of reading ±0.02 % of range
Up to 1 kHz±0.03 % of reading ±0.02 % of range
Up to 10 kHz±0.1 % of reading ±0.05 % of range
Up to 100 kHz±4 % of reading ±0.1 % of range
Up to 1000 kHz±5 % of reading ±0.5 % of range
Noise floor, Typ.Sample rate / Range →2000 V1000 V400 V200 V
15 MS/s-88 dB-87 dB-85 dB-81 dB
1 MS/s-97 dB-94 dB-89 dB-83 dB
100 kS/s-108 dB-107 dB-101 dB-96 dB
10 kS/s-118 dB-117 dB-111 dB-106 dB
CMR, Typ. (Min.)101 dB (88 dB) @ 50 Hz, 75 dB (68 dB) @ 400 Hz
Gain DriftTypical 20 ppm/K, max. 40 ppm/K
Offset DriftTypical 1.5 mV/K + 1 ppm of range/K, max 3 mV/K + 2 ppm of range/K
Gain Linearity< 0.01 %
Input CouplingDC
Input Impedance10 MΩ | 1 pF
Current Mode
Ranges±2000 A, ±1000 A, ±400 A, ±200 A
ADC TypeHybridADC - alias free up to 2 MS/s, 16-bit up to 15 MS/s
Sampling RateSimultaneous 15 MS/s
FilteringAAF 1 MHz (6th order)
Analogue bandwidth (-3 dB)> 500 kHz
Working current (continuous)±1000 ADC
Measuring current±2000 ADC
Maximum withstand peak currentmin. -1700 A, max. 2000 A (see 2)
Primary / Secondary Ratio1 : 1680
Input Accuracy (Ip @ 10 Arms)Signal frequencyAccuracy
DC±0.1 % of reading ±50 ppm of range ±0.05 A
Up to 10 kHz±0.3 % of reading
Up to 20 kHz±0.8 % of reading
Up to 200 kHz±4 % of reading
Gain drift (current)typ. 40 ppm/K
Offset drifttyp. 200 uA/K
Linearity error @ 1000 A rangetyp. 50 ppm
Hysteresistyp. 50 uA/A
Flatness DC - 50 Hz Overall accuracy @ 25°Ctyp. 5 mdB
Flatness 50 Hz - 1 kHz Overall accuracy @ 25°Ctyp. 20 mdB
HV harness
Cable size (single core)120 mm295 mm270 mm250 mm235 mm2
Rated DC current344 A292 A245 A198 A158 A
Cable glandM32M32M32M25M25
General specifications
Power
Power Supply9 - 48 V DC (see 3) PWR + DATA + SYNC (Lemo 1T 8-pin)
Power consumptionTyp. 13 W (Max. 23 W)
Environmental
Operating Temperature-20 °C to 70 °C
Storage Temperature-40 °C to 85 °C
IP ratingIP67
Shock & VibrationEN 60068-2-6:2008 Environmental testing -- Part 2-6: Tests - Test Fc: Vibration (sinusoidal) EN 60068-2-27:2009 Environmental testing -- Part 2-27: Tests - Test Ea and guidance: Shock
Interfaces
EthernetGbE (XCP, OPC UA) incl. IEEE1588v2 synchronization (PTP) (LEMO 1T 8-pin)
CANCAN 2.0 (DSUB-9)
SynchronizationIEEE1588v2 synchronization (PTP) (LEMO 1T 8-pin), IRIG (D-SUB9)
High-Voltage InterlockHW ready (chassis covers + cables)
Additional Specifications
GroundingM6 insert
Dimensions245 x 151 x 63 mm
Weight (incl. cables and connectors)4.4 kg (max. 7 kg)
Rev: 1663912800
1) ±2000 V is measuring range. Device is designed according to IEC 61010-1. Maximum conitnuous working voltage is 1000 VDC or 1000 VAC.
2) ±2000 A is measuring range. Unit can withstand peak value of 2000 A for 100 ms with 1 s recovery time.
3) At the Power Supply lower than 12 V, the maximum measuring current is limited to ±850 A

SIRIUSi-XHS-PWR-1xHV-1xDCCT-1000A-CON tehnične specifikacije

SIRIUSi-XHS-PWR specificationsSIRIUSi-XHS-PWR-1xHV-1xDC-CT-1000A-CON
Input typesVoltage, Current (single phase)
Voltage Mode
Voltage ranges±2000 V, ±1000 V, ±400 V, ±200 V
ADC TypeHybrid ADC - alias free up to 2 MS/s, 16-bit up to 15 MS/s
Sampling RateSimultaneous 15 MS/s
FilteringAAF 1 MHz (6th order)
Analog bandwidth (-3 dB)5 MHz
Working voltage (continuous)1000 VDC or 1000 VAC
Measuring voltage (max.)±2000 V (see 1)
Input AccuracySignal frequencyAccuracy
DC±0.03 % of reading ±0.02 % of range
Up to 1 kHz±0.03 % of reading ±0.02 % of range
Up to 10 kHz±0.1 % of reading ±0.05 % of range
Up to 100 kHz±4 % of reading ±0.1 % of range
Up to 1000 kHz±5 % of reading ±0.5 % of range
Noise floor, Typ.Sample rate / Range →2000 V1000 V400 V200 V
15 MS/s-88 dB-87 dB-85 dB-81 dB
1 MS/s-97 dB-94 dB-89 dB-83 dB
100 kS/s-108 dB-107 dB-101 dB-96 dB
10 kS/s-118 dB-117 dB-111 dB-106 dB
CMR, Typ. (Min.)101 dB (88 dB) @ 50 Hz, 75 dB (68 dB) @ 400 Hz
Gain DriftTypical 20 ppm/K, max. 40 ppm/K
Offset DriftTypical 1.5 mV/K + 1 ppm of range/K, max 3 mV/K + 2 ppm of range/K
Gain Linearity< 0.01 %
Input CouplingDC
Input Impedance10 MΩ | 1 pF
Current Mode
Current ranges±2000 A, ±1000 A, ±400 A, ±200 A
ADC TypeHybridADC - alias free up to 2 MS/s, 16-bit up to 15 MS/s
Sampling RateSimultaneous 15 MS/s
FilteringAAF 1 MHz (6th order)
Analogue bandwidth (-3 dB)> 500 kHz
Working current (continuous)±1000 ADC
Measuring current±2000 ADC
Maximum withstand peak currentmin. -1700 A, max. 2000 A (see 2)
Primary / Secondary Ratio1 : 1680
Input Accuracy (Ip @ 10 Arms)Signal frequencyAccuracy
DC±0.1 % of reading ±50 ppm of range ±0.05 A
Up to 10 kHz±0.3 % of reading
Up to 20 kHz±0.8 % of reading
Up to 200 kHz±4 % of reading
Gain drift (current)typ. 40 ppm/K
Offset drifttyp. 200 uA/K
Linearity error @ 1000 A rangetyp. 50 ppm
Hysteresistyp. 50 uA/A
Flatness DC - 50 Hz Overall accuracy @ 25°Ctyp. 5 mdB
Flatness 50 Hz - 1 kHz Overall accuracy @ 25°Ctyp. 20 mdB
Connectors
HV ConnectorsPowerLok 500 Series
Device connectorsPL00X-501-10D10-2 (+), PL00Y-501-10D10-2 (-)
Mating connectors (cable)PL18X-501-?-2-5 (+), PL18Y-501-?-2-5 (-) (see3)
General specifications
Power
Power Supply9 - 48 V DC (see 4) PWR + DATA + SYNC (Lemo 1T 8-pin)
Power consumptionTyp. 13 W (Max. 23 W)
Environmental
Operating Temperature-20 °C to 70 °C
Storage Temperature-40 °C to 85 °C
IP ratingIP67
Shock & VibrationEN 60068-2-6:2008 Environmental testing -- Part 2-6: Tests - Test Fc: Vibration (sinusoidal) EN 60068-2-27:2009 Environmental testing -- Part 2-27: Tests - Test Ea and guidance: Shock
Interfaces
EthernetGbE (XCP, OPC UA) incl. IEEE1588v2 synchronization (PTP) (LEMO 1T 8-pin)
CANCAN 2.0 (DSUB-9)
SynchronizationIEEE1588v2 synchronization (PTP) (LEMO 1T 8-pin), IRIG (D-SUB9)
High-Voltage InterlockHW ready (chassis covers + cables)
Additional Specifications
GroundingM6 insert
Dimensions245 x 151 x 63 mm
Weight (incl. cables and connectors)4.4 kg
Rev: 1663912800
1) ±2000 V is measuring range. Device is designed according to IEC 61010-1. Maximum conitnuous working voltage is 1000 VDC or 1000 VAC.
2) ±2000 A is measuring range. Unit can withstand peak value of 2000 A for 100 ms with 1 s recovery time.
3) ? denotes HV cable size, e.g. PL18X-501-95-2-5 is designed for 95mm2 cable cross-section
4) At the Power Supply lower than 12 V, the maximum measuring current is limited to ±850 A

Tehnične specifikacije SIRIUSi-XHS-PWR-1xHV-1xDCCT-250A-UNI

SIRIUSi-XHS-PWR specificationsSIRIUSi-XHS-PWR-1xHV-1xDC-CT-250A-UNI
Input typesVoltage, Current (single phase)
Voltage Mode
Voltage ranges±2000 V, ±1000 V, ±400 V, ±200 V
ADC TypeHybrid ADC - alias free up to 2 MS/s, 16-bit up to 15 MS/s
Sampling RateSimultaneous 15 MS/s
FilteringAAF 1 MHz (6th order)
Analog bandwidth (-3 dB)5 MHz
Working voltage (continuous)1000 VDC or 1000 VAC
Measuring voltage (max.)±2000 V (see 1)
Input AccuracySignal frequencyAccuracy
DC±0.03 % of reading ±0.02 % of range
Up to 1kHz±0.03 % of reading ±0.02 % of range
Up to 10 kHz±0.1 % of reading ±0.05 % of range
Up to 100 kHz±4 % of reading ±0.1 % of range
Up to 1000 kHz±5 % of reading ±0.5 % of range
Noise floor, Typ.Sample rate / Range →2000 V1000 V400 V200 V
15 MS/s-88 dB-87 dB-85 dB-81 dB
1 MS/s-97 dB-94 dB-89 dB-83 dB
100 kS/s-108 dB-107 dB-101 dB-96 dB
10 kS/s-118 dB-117 dB-111 dB-106 dB
CMR, Typ. (Min.)101 dB (88 dB) @ 50 Hz, 75 dB (68 dB) @ 400 Hz
Gain DriftTypical 20 ppm/K, max. 40 ppm/K
Offset DriftTypical 1.5 mV/K + 1 ppm of range/K, max 3 mV/K + 2 ppm of range/K
Gain Linearity< 0.01 %
Input CouplingDC
Input Impedance10 MΩ | 1 pF
Current Mode
Current ranges±500 A, ±250 A, ±100 A, ±50 A
ADC TypeHybridADC - alias free up to 2 MS/s, 16-bit up to 15 MS/s
Sampling RateSimultaneous 15 MS/s
FilteringAAF 1 MHz (6th order)
Analogue bandwidth (-3 dB)> 500 kHz
Working current (continuous)Limited with the rated current of HV harness (see 2)
Measuring current±500 ADC
Maximum withstand peak currentmin. -500 A, max. 500 A
Primary / Secondary Ratio1 : 1680
Input Accuracy (Ip @ 10 Arms)Signal frequencyAccuracy
DC±0.1 % of reading ±50 ppm of range ±0.05 A
Up to 10kHz±0.3 % of reading
Up to 20 kHz±0.8 % of reading
Up to 200kHz±4 % of reading
Gain drift (current)typ. 40ppm/K
Offset drifttyp. 200 uA/K
Linearity error @ 250 A rangetyp. 50 ppm
Hysteresistyp. 50 uA/A
Flatness DC - 50 Hz Overall accuracy @ 25°Ctyp. 5 mdB
Flatness 50 Hz - 1 kHz Overall accuracy @ 25°Ctyp. 20 mdB
Cable size (multi core)6 mm2
Rated DC current54 A
General specifications
Power
Power Supply9 - 48 V DC PWR + DATA + SYNC (Lemo 1T 8-pin)
Power consumptionTyp. 13 W (Max. 20 W)
Environmental
Operating Temperature-20 to 70 °C
Storage Temperature-40 to 85 °C
IP ratingIP67
Shock & VibrationEN 60068-2-6:2008 Environmental testing -- Part 2-6: Tests - Test Fc: Vibration (sinusoidal) EN 60068-2-27:2009 Environmental testing -- Part 2-27: Tests - Test Ea and guidance: Shock
Interfaces
EthernetGbE (XCP, OPC UA) incl. IEEE1588v2 synchronization (PTP) (LEMO 1T 8-pin)
CANCAN 2.0 (DSUB9)
SynchronizationIEEE1588v2 synchronization (PTP) (LEMO 1T 8-pin), IRIG (D-SUB9)
High-Voltage InterlockYES (chassis covers + cables)
Additional Specifications
GroundingM6 insert
Dimensions245 x 109 x 63 mm
Weight2.3 kg
Rev: 1667217600
1) ±2000 V is measuring range. Device is designed according to IEC 61010-1. Maximum conitnuous working voltage is 1000 VDC or 1000 VAC.
2) ±500 A is measuring range. Maximum current is limited with specified rated current of HV harness (connectors and cables).

Specifikacije analize moči

FunctionalityDewesoft Power Analyzer
Power Analysis for DC and AC
Power Analysis PP, Q, S, PF, cos phi, D (Distortion), DH (Harmonic distortion), QH (reactive power of harmonics) (for each phase and total)
Fundamental PowerP_H1, Q_H1, S_H1, cos phi_H1, phi_H1 (for each phase and total)
Voltage and CurrentRMS, RM (star and delta)
Energy CalculationTotal, positive and negative (e.g. Recuperation)
Efficiency
Wiring SchematicsDC, 1-phase, 2-phase, 3-phase delta, 3-phase star, 3-phase V, 3-phase Aron, 3-phase two meters, 6-phase (R2, R4, R8), 7-phase (R2, R4, R8, 12-phase (R4, R8))
Star-Delta Calculation✓ (waveform and RMS values)
Frequencies16.7 Hz, 25 Hz, 50 Hz, 60 Hz, 400 Hz, 800 Hz, Variable from 10 Hz up to 3 kHz
Frequency SourceVoltage, current, external
Period ValuesU, I, P, Q, S, symmetrical components for 1⁄2, 1, 2 or 4 periods and selectable Overlap up to 99 %
Power AveragingSelectable - starting from 1 ms , Multiple Averaging (e.g. 20 ms, 60 s, 600 s) possible

Specifikacije za analizo kakovosti električne energije

FunctionalityDewesoft Power Analyzer
Harmonics (according to IEC61000-4-7)up to 150 kHz for voltage, current, active-, reactive power, phase angle and impedance
Variable Sidebands and Half Sidebands (according to IEC61000-4-7)
Harmonic Smoothing Filter (according to IEC61000-4-7)
Interharmonics (according to IEC61000-4-7)
Total Harmonic Distortion (THD) (according to IEC61000-4-7) Voltage and current (Total, odd and even) - selectable up to 150 kHz
Total Interharmonic Distortion (TIHD) and K-factor (according to IEC61000-4-7)Voltage and current (Total, odd and even) - selectable up to 150 kHz
Higher Frequencies (according to IEC61000-4-7) up to 150 kHz (grouping in 200 Hz and 2 kHz bands)
Flicker (according to IEC61000-4-15)selectable PST and PLT
Flicker Emission (according to IEC61400-21)
Rapid Voltage Changes (according to IEC61000-4-15)selectable steady state and hysteresis
Symmetrical Components (according to IEC61000-4-30)Zero-, positive- & negative system for voltage and current (absolute or relative to fundamental)
Additional Symmetrical Components (according to IEC61400-21) Active and reactive parts for zero-, positive- & negative system

Specifikacije programske opreme analizatorja moči

Software specifications
FunctionalityDewesoft Power Analyzer
Power Analysis
Power Quality Analysis
Database Storing
Post Processing
Math Library
Data logging - Raw data storing✓ (data Storing at Full Sampling rate up to 15 MS/s per channel)
Scope✓ (up to 8 graphs in one diagram, Zoom In- and Out)
Vector Scope✓ (1-, 2-, 3-phase systems, Clarke transform, Park transform)
FFT✓ (up to 1⁄2 of Sampling Rate)
Harmonic FFT✓ (up to 1⁄4 of Sampling Rate)
Transient Recording✓ (up to 15 MS/s)
Triggering ChannelsAnalog, Digital, Counter, Math, Power, etc.
Triggering optionsSimple edge (rising, falling), Window (two-levels: entering, leaving), Pulsewidth
(longer or shorter than duration), Window and Pulsewidth, Slope Trigger (rising or
falling slope with steepness)
Motor Efficiency Analysis✓ (available as an upgrade)
Efficiency mapping

Natančnost frekvence

Line frequencyAccuracy
16.67 ... 60 Hz0.0005 % of reading
400 Hz0.01 % of reading
800 Hz0.02 % of reading
Variable0.05 % of reading
Condition: SR > 50 * fin