SIRIUS® XHS 4xHV, 4xLV Spécifications techniques

Spécifications de l'analyse de puissance

FunctionalityDewesoft Power Analyzer
Power Analysis for DC and AC
Power Analysis PP, Q, S, PF, cos phi, D (Distortion), DH (Harmonic distortion), QH (reactive power of harmonics) (for each phase and total)
Fundamental PowerP_H1, Q_H1, S_H1, cos phi_H1, phi_H1 (for each phase and total)
Voltage and CurrentRMS, RM (star and delta)
Energy CalculationTotal, positive and negative (e.g. Recuperation)
Efficiency
Wiring SchematicsDC, 1-phase, 2-phase, 3-phase delta, 3-phase star, 3-phase V, 3-phase Aron, 3-phase two meters, 6-phase (R2, R4, R8), 7-phase (R2, R4, R8, 12-phase (R4, R8))
Star-Delta Calculation✓ (waveform and RMS values)
Frequencies16.7 Hz, 25 Hz, 50 Hz, 60 Hz, 400 Hz, 800 Hz, Variable from 10 Hz up to 3 kHz
Frequency SourceVoltage, current, external
Period ValuesU, I, P, Q, S, symmetrical components for 1⁄2, 1, 2 or 4 periods and selectable Overlap up to 99 %
Power AveragingSelectable - starting from 1 ms , Multiple Averaging (e.g. 20 ms, 60 s, 600 s) possible

Spécifications pour l'analyse de la qualité de l'énergie

FunctionalityDewesoft Power Analyzer
Harmonics (according to IEC61000-4-7)up to 150 kHz for voltage, current, active-, reactive power, phase angle and impedance
Variable Sidebands and Half Sidebands (according to IEC61000-4-7)
Harmonic Smoothing Filter (according to IEC61000-4-7)
Interharmonics (according to IEC61000-4-7)
Total Harmonic Distortion (THD) (according to IEC61000-4-7) Voltage and current (Total, odd and even) - selectable up to 150 kHz
Total Interharmonic Distortion (TIHD) and K-factor (according to IEC61000-4-7)Voltage and current (Total, odd and even) - selectable up to 150 kHz
Higher Frequencies (according to IEC61000-4-7) up to 150 kHz (grouping in 200 Hz and 2 kHz bands)
Flicker (according to IEC61000-4-15)selectable PST and PLT
Flicker Emission (according to IEC61400-21)
Rapid Voltage Changes (according to IEC61000-4-15)selectable steady state and hysteresis
Symmetrical Components (according to IEC61000-4-30)Zero-, positive- & negative system for voltage and current (absolute or relative to fundamental)
Additional Symmetrical Components (according to IEC61400-21) Active and reactive parts for zero-, positive- & negative system

Spécifications du logiciel de l'analyseur de puissance

Software specifications
FunctionalityDewesoft Power Analyzer
Power Analysis
Power Quality Analysis
Database Storing
Post Processing
Math Library
Data logging - Raw data storing✓ (data Storing at Full Sampling rate up to 15 MS/s per channel)
Scope✓ (up to 8 graphs in one diagram, Zoom In- and Out)
Vector Scope✓ (1-, 2-, 3-phase systems, Clarke transform, Park transform)
FFT✓ (up to 1⁄2 of Sampling Rate)
Harmonic FFT✓ (up to 1⁄4 of Sampling Rate)
Transient Recording✓ (up to 15 MS/s)
Triggering ChannelsAnalog, Digital, Counter, Math, Power, etc.
Triggering optionsSimple edge (rising, falling), Window (two-levels: entering, leaving), Pulsewidth
(longer or shorter than duration), Window and Pulsewidth, Slope Trigger (rising or
falling slope with steepness)
Motor Efficiency Analysis✓ (available as an upgrade)
Efficiency mapping

Précision de la fréquence

Line frequencyAccuracy
16.67 ... 60 Hz0.0005 % of reading
400 Hz0.01 % of reading
800 Hz0.02 % of reading
Variable0.05 % of reading
Condition: SR > 50 * fin

Spécifications du système SIRIUS® XHS

Power specs
Power Supply9 - 48 V DC
Power consumptionTyp. 30 W
Environmental
Operating Temperature-10 to 50 °C
Storage Temperature-40 to 85 °C
Humidity5 to 95 % RH non condensing @ 60 °C
IP ratingIP20
Shock & VibrationShock (EN 60068-2-27:2009) 75 g, 6 ms, half-sine (25x pos./neg in each axis) Random Vibration (EN 60721-3-2: 1997 - Class 2M2) Sweep sinus Vibration (EN 60068-2-6:2008)
Dimensions266 mm x 149 mm x 55 mm 266 mm x 149 mm x 69 mm (extended slice)
Interfaces
Ethernet1 GbE (XCP, OPC UA) incl. IEEE1588v2 synchronization (PTP) (RJ45)
USBUSB 3.0 (Type C)
CANCAN 2.0, CAN-FD (DSUB9)
Sync Input/Ouput
Level (Input/Output)TTL compatible
Max. Output Current±24 mA (±50 mA for 1 sec)
Synchronization
Delay between slices± 1 μs ( to the reference clock )
Max. Sync-cable length100 m (Master/Slave), 200 m (IRIG)

Spécifications du SIRIUS® XHS HybridADC

ADC
TypeHybrid ADC
High bandwidth mode16 bit SAR, up to 15 MS/s
Alias free mode24 bit Sigma Delta, up to 2 MS/s
Digital low pass filterSoftware selectable (off, alias free)
Analog HW filterAnalog 1 MHz 6th order (Butterworth)
Alias free filter0.5*fs cutoff, 0.45*fs pass band, 0.55*fs stop band to -100dB, zero phase distortion, available below 2MS/s

Spécification du filtrage du SIRIUS® XHS HybridADC

Filter ModeHigh-bandwidth1 MHz LPFHigh-dynamic
Max. SR15 MS/s15 MS/s2 MS/s
SW settingsOff1 MHzAAF
Bandwidth (-3 dB)5 MHz1 MHz0.49 fs
Passband frequency (-0.1 dB)500 kHz500 kHz0.45 fs
Passband flatness± 0.1 dB± 0.1 dB± 0.1 dB
Alias-free bandwidth7.5 MHz0.45 fs
HW filter
Stopband frequency7.5 MHz0.55 fs
Stopband attenuation-100 dB-100 dB

Spécifications des compteurs numériques SIRIUS® XHS

Counter Input
Input Channels (See 1)8
ModesEvent counting, waveform timing, encoder, tacho, geartooth,...
Counter timebase150 MHz
Time base accuracyTypical: 0.5 ppm, Max: 2 ppm
Max. Bandwidth15 MHz
Max. Sampling Rate5 MS/s
Input FilterOff, 500 ns, 1 μs, 2 μs, 4 μs, 5 μs and 7.5 μs
Counter resolution32-bit
Input Compatibility, Level3.3 V TTL/CMOS (Low: <0.8 V, High >2 V)
Input termination, impedance100 kΩ pull-up to +3.3 V
Input Protection±25 V continuous
Digital Output
Output Channels8 connectors, 1 line per connector
Output5 V digital output
Maximum current32 mA
Output series resistance50 Ω
Sensor Power Supply
Output Voltage +5 V5 V +/- 10 % 500 mA typ. Output short protected (Shared between 8 ports)
Output Voltage +12 V (See 2)12 V +/-10 % 300 mA typ. Output short protected (Shared between 8 ports)
1) Digital inputs on Counter (individual A, B, Z input) can be used only for Trigger condition 2) Maximum 2x DS-TACHO4 laser probes supported.
Rev: 1702296000

Aperçu des amplificateurs SIRIUS® XHS

XHS-HVXHS-LVXHS-ACC
ConnectorsBANANADB9, BNC, BANANABNC
Channels per slice888
Data rate / channel15 MS/s15 MS/s15 MS/s
Resolution16-bit (24-bit @ 1 MS/s)16-bit (24-bit @ 1 MS/s)16-bit (24-bit @ 1 MS/s)
Bandwidth5 MHz5 MHz5 MHz
Voltage ranges±2000 V ... ±200 V±100 V ... ±50 mV±10 V ... ±200 mV
Input couplingDCDC, AC 1 HzDC, AC 0.1 Hz, AC 1 Hz
Sensor excitation2.5..30 V bipolar, 2..24 V unipolar, max. 0.2 A / 2 WIEPE 2 mA, 4 mA, 8 mA, 12 mA, 16 mA, 20 mA
Bridge connection
Programmable shunt
IEPE inputDSI-ACC
Resistance
Temperature (PTx)DSI-RTD
ThermocoupleDSI-TH
Potentiometer
LVDTDSI-LVDT
ChargeDSI-CHG
Currentext. shunt DSI20mA, DSI5Aext. shunt
TEDS
Isolation voltageCATII 1000 V420 Vdc (300 Vacrms)140 Vdc (100 Vacrms)
Power consumption per channel (typ.)1.45 W/ch1.55 - 5 W/ch1.8 W/ch
Advanced functionsHigh voltage, high bandwidth, high isolationHigh sensor excitation and multi rangeSensor error detection, high speed
Rev: 1701403200

Spécifications de l'amplificateur SIRIUS® XHS HV

Inputs
ADC TypeHybrid ADC - alias free up to 2 MS/s, 16-bit up to 15 MS/s
Sampling RateSimultaneous 15 MS/s
FilteringAAF 1 MHz (6th order)
Analog bandwidth (-3 dB)5 MHz
Voltage Mode
Ranges±2000 V, ±1000 V, ±400 V, ±200 V
Input AccuracySignal frequencyAccuracy
DC to 1 kHz±0.03 % of reading ±0.02 % of range
Up to 10 kHz±0.1 % of reading ±0.05 % of range
Up to 100 kHz±4 % of reading ±0.1 % of range
Up to 1000 kHz±5 % of reading ±0.5 % of range
Noise floor, Typ.Sample rate / Range →2000 V1000 V400 V200 V
15 MS/s-84 dB-82 dB-79 dB-75 dB
1 MS/s-95 dB-93 dB-85 dB-78 dB
100 kS/s-108 dB-106 dB-100 dB-95 dB
10 kS/s-117 dB-115 dB-109 dB-104 dB
CMR, Typ.101 dB @ 50 Hz, 74 dB @ 400 Hz
Channel Crosstalk-116 dB typ. @ 50 Hz, -89 dB typ. @ 1 kHz
Gain DriftTypical 20 ppm/K, max. 40 ppm/K
Offset DriftTypical 1.5 mV/K + 1 ppm of range/K, max 3 mV/K + 2 ppm of range/K
Gain Linearity< 0.01 %
Input CouplingDC
Input Impedance10 MΩ | 1 pF
Overvoltage categoryCAT II 1000 V
Maximum input voltage2000 Vdc or 2000 Vpk (1414 Vacrms)
Additional Specifications
Input connectorBanana
Rev: 1698322800

Spécifications de l'amplificateur SIRIUS® XHS LV

InputsVoltage
ADC TypeHybrid ADC - alias free up to 2 MS/s, 16-bit up to 15 MS/s
Sampling RateSimultaneous 15 MS/s
FilteringAAF 1 MHz (6th order)
Analog bandwidth (-3 dB)5 MHz
Voltage Mode
Ranges±100, 50, 20, 10, 5, 2.5, 1, 0.5, 0.25, 0.1 and 0.05 V
Input AccuracyDC ±0.03 % of reading ±0.02 % of range ±100 μV (2 mV for Ranges ≥ 10 V)
Up to 1 kHz±0.03 % of reading ±0.02 % of range
Up to 10 kHz±0.1 % of reading ±0.05 % of range
Up to 100 kHz±2 % of reading ±0.1 % of range
Up to 1000 kHz±5 % of reading ±0.5 % of range
Noise floor, Typ.Sample rate, filter \ Range →100 V5 V0.5 V0.05 V
15 MS/s, Filter off-82 dB-78 dB-65 dB-51 dB
1 MS/s, Filter on-94 dB-95 dB-88 dB-69 dB
100 kS/s, Filter on-104 dB-105 dB-100 dB-81 dB
10 kS/s, Filter on-113 dB-114 dB-107 dB-87 dB
CMR, Typ. Frequency \ Range →100 V5 V0.5 V0.05 V
50 Hz91 dB102 dB98 dB100 dB
400 Hz86 dB75 dB90 dB98 dB
10 kHz66 dB74 dB82 dB83 dB
Channel CrosstalkRange ≥ 10 V: -105 dB @ 10 kHz Range < 10 V: -120 dB @ 10 kHz
Gain DriftTypical 10 ppm/K, max. 30 ppm/K
Offset DriftRange ≥ 10 V: Typ. 100 μV/K + 1 ppm of range/K, Max 250 μV/K + 2 ppm of range/K Range < 10 V: Typ. 4 μV/K + 1 ppm of range/K, Max 8 μV/K + 2 ppm of range/K
Gain Linearity< 0.03 %
Input Coupling (see 1)DC, AC 1 Hz
Input ImpedanceRange < 10 V: 10 MΩ; Range ≥ 10 V: 1 MΩ | 110 pF between INx to GND
Overvoltage ProtectionRange ≥ 10 V: 300 V continuous Range < 10 V: 100 V (200 V peak for 10 msec)
Max. common mode voltage Isolation voltage420 Vdc (300 Vacrms)
Excitation Voltage
PolarityUnipolar or Bipolar Software selectable (programmable 16-bit DAC)
Excitation LevelUnipolar 1 .. 24 V; Predefined levels: 1, 2, 2.5, 5, 10, 12, 15, 20 and 24 VDC
Excitation LevelBipolar 2.5 .. 30 V; Predefined levels: 2.5, 5, 10, 12, 15, 24 and 30 VDC
Drift±10 ppm/K ±150 μV/K
Stability 10% to 90% load (bipolar)< 0.01 %
Current limit200 mA (2 Watt max. per Channel, 12 Watt max. per Slice)
Additional Specifications
Input connectorDSUB 9, BNC, Banana (others on request)
TEDS supportStandard + DSI® adapters
Rev: 1701403200
1) In- must be within ±10 V referred to GND (iso); for Ranges ≥ 10 V the DC value of In- is not rejected
Haut de page